| SynchroSonic
®
U/HVG50 Specifications |
|
SynchroSonic U/HVG50
Combination Ultrasound with optional three shelf utility cart
Input Power Requirements
| Line Voltage |
120 Vac, 60
Hz
(Special voltages available
on request)
|
Ultrasonic Generator
(Complies with FDA Regulation
21 CFR 1050.10 and FCC Regulations)
| Waveform |
amplitude modulated
at 120 Hz ± 1%
|
| Output Mode |
adjustable
interrupted & non-interrupted
|
| Output Power |
0 to 20 W total
± 20%
0 to 5 W/cm² ±
20%
|
| Temporal Peak/Temporal Average
Intensity |
2.8:1 ±
20%
|
| Treatment Timer |
adjustable
0 to 30 minutes
|
Applicator (Transducer)
| Crystal Material |
barium-titanate
|
| Applicator Beam Type |
collimating
|
| ERA (Effective Radiating
Area) |
4 cm² ±
20% |
| BNR (Beam Nonuniformity
Ratio) |
5.5:1 maximum
|
| Small Faceplate Size |
1.5" diameter
|
| Standard Faceplate Size |
2.25" diameter
|
Caution: Federal law (USA)
restricts this device to sale by or on the order of a licensed practitioner |
Ultrasound Features
-
Lightweight standard transducer
-
Optional QuickConnect™ transducer
-
Optional small transducer with
QuickConnect cable
-
Cable fault alarm system
-
Transducer protection switch
-
Adjustable interrupted output
mode
-
0 to 20 watt output •
30 minute timer
High Volt Stimulator
Features
-
Dual channel, 4 pad high volt
pulsed dc stimulator
-
Selectable frequencies - pulsation
to tetanize
-
Selectable surge and reciprocal
rates - ramped
-
Intensity reset circuit •
Combination therapy capability
Low Voltage Galvanic
Features
-
Single channel, 2 pad low voltage
galvanic (dc)
-
0 to 10 mA current for iontophoresis
treatments
-
Adjustable surge and alternating
polarity surge rates
-
Combination: iontophoresis/phonophoresis
with FDA approved substances
-
Two Year Warranty - Generator
- Parts and Labor
-
One Year Warranty - Transducer
- Parts and Labor
-
CSA Approved
Electrical Generator - High
Volt
| Waveform |
dc "needle"
pulses with no parasitic spike
|
| Output Voltage |
0 to 500 peak
V
|
| Output Intensity |
20 µC
maximum charge
per pulse into
100 Ohm load
|
| Frequency Range |
1 to 160 pulses
per s
|
| Reciprocal/Surge Duration |
2.5, 5 and
10 s
|
Electrical Generator - Low
Voltage Galvanic
| Description |
variable constant
current source dc output
|
| Output Intensity |
0 to 10 mA
|
| Surge Duration |
variable 3.33
s on/off to 0.50 s on/off
|
| Alternating Polarity Surge |
variable 3.33
± to 0.50 s ±
|
Instrument U/HVG50
| Overall Dimensions |
19½"
W x 11" D x 7" H
|
| Weight (including transducer) |
18 lbs.
|
Ordering Information U/HVG50
| Standard Soundhead |
32-U/HVG50
|
| Standard QuickConnect Soundhead |
32-U/HVG50-D
|
| Small QuickConnect Soundhead |
32-U/HVG50-SM
|
| Standard/Small QuickConnect
Soundhead |
32-U/HVG50-D2
|
|
Federal law
(USA) restricts Amrex electrotherapy instruments to sale by
or on the order of a licensed practitioner
\\ Welcome
|| Dealers
|| Products
|| Instructional Videos
|| Accessories //
\\ ATS
|| L.A. Door to Door Factory Service
|| Patient Leads
|| Trade-In Program //
\\ News
|| Amrex 1-3.3 MHz Ultrasound Output
|| Contact Us //
\\ User's Guides
|| History //

a division of Amrex-Zetron, Inc.
641 East Walnut Street, Carson, California
90746
Telephone: (800) 221-9069 or (310)
527-6868
Fax: (310) 366-7343
Web Site: http://www.amrex-zetron.com

Copyright © 1998 Amrex-Zetron
Corporation. All rights reserved.
Last modified on January 22,
2007
Web Site Creator:
Produced and Maintained by Paul Elliott: Designer/Graphics
E-Mail: pkelliot@gte.net
Comments, suggestions welcome!